Title: Comparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selection
Authors: Sabong Srivannaboon
Chareenun Chirapapaisan
Pratuangsri Chonpimai
Sunisa Koodkaew
Mahidol University
Keywords: Medicine
Issue Date: 1-Jan-2015
Citation: Journal of Cataract and Refractive Surgery. Vol.41, No.2 (2015), 364-371
Abstract: © 2015 ASCRS and ESCRS. Purpose To compare corneal astigmatism measurements obtained using a new optical biometer (AL-Scan) and a standard biometer based on partial coherence interferometry (IOLMaster 500) for toric intraocular lens (IOL) selection. Setting Siriraj Hospital, Mahidol University, Bangkok, Thailand. Design Prospective comparative study. Methods Keratometry (K) readings were taken with the 2 biometers in eyes of cataract patients. The corneal astigmatism readings from the 2 devices were compared. The cylinder measurements were transformed into J0 and J45 vector components for analysis, and the devices' values were compared using the paired t test. Correlation coefficients were calculated. The Bland-Altman method was used to evaluate the agreement between the sets of measurements. Results One hundred thirty-seven eyes (81 patients) were enrolled. The magnitude of astigmatism between the new biometer 2.4 mm zone and the standard biometer was similar (P =.19). There was a statistically significant difference between values from the new biometer 3.3 mm zone and the standard biometer (P <.05). There was good correlation in the overall magnitude of astigmatism between devices. There was a trend of low correlation toward the flat cornea (K <43.0 diopters [D]), low corneal cylinder (<1.0 D), and with-the-rule astigmatism. The IOL toricity calculated from both devices was similar. Conclusions Corneal astigmatism measurements with the 2 biometers were similar in magnitude, especially in 2.4 mm zone of the new biometer. There was no difference in the toric IOL selections between the devices. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.
URI: https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84923095888&origin=inward
http://repository.li.mahidol.ac.th/dspace/handle/123456789/36757
ISSN: 18734502
08863350
Appears in Collections:Scopus 2011-2015

Files in This Item:
There are no files associated with this item.


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.