Thin Pa Pa AyeZwe Lin HtetThamvarit SinghavilaiPhornphop NaiyanetrMahidol University2018-11-232018-11-232015-11-04Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS. Vol.2015-November, (2015), 3303-33061557170X2-s2.0-84953259305https://repository.li.mahidol.ac.th/handle/20.500.14594/35793© 2015 IEEE. Intra-aortic balloon pump (IABP) has been used in clinical treatment as a mechanical circulatory support device for patients with heart failure. A computer model is used to study the effect on coronary blood flow (CBF) with different balloon cycles under both normal and pathological conditions. The model of cardiovascular and IABP is developed by using MATLAB SIMULINK. The effect on coronary blood flow has been studied under both normal and pathological conditions using different balloon cycles (balloon off; 1/4; 1/2; 1/1). A pathological heart is implemented by reducing the left ventricular contractility. The result of this study shows that the rate of balloon cycles is related to the level of coronary blood flow.Mahidol UniversityComputer ScienceEngineeringMedicineEffect of intra-aortic balloon pump on coronary blood flow during different balloon cycles support A computer studyConference PaperSCOPUS10.1109/EMBC.2015.7319098