Browsing by Author "Christian Breukers"
Now showing 1 - 1 of 1
- Results Per Page
- Sort Options
Publication Metadata only Clinical evaluation of a simple image cytometer for CD4 enumeration on hiv-infected patients(2010-01-28) Xiao Li; Christian Breukers; Aurel Ymeti; Kovit Pattanapanyasat; Kasama Sukapirom; Leon W M M Terstappen; Jan Greve; University of Twente; Mahidol UniversityBackground: Affordable, easy-to-use, and reliable CD4+ T lymphocyte enumeration systems are needed in resource-constrained settings to monitor HIV. Methods: A simple image cytometer was used to count fluorescently labeled CD4+ T and CD8+ T lymphocytes from CD3 immunomagnetically selected cells on blood specimens of 460 HIV-1-infected patients in Siriraj Hospital, Bangkok, Thailand. Results were compared with flow cytometry (FCM). Results: CD4+ T lymphocyte counts by image cytometer were comparable (R ≥ 0.97) with those by the FACSCount and the FACScan with a bias of 7.3 and 9.1%, respectively. At very low CD4+ T lymphocyte counts (≤50/μl) some over-count outliers were observed by the FACScan and image cytometer when compared with the FACSCount. For CD8 enumeration, the image cytometer showed a good correlation (R 5 0.96) and a consistent undercount (-17%) when compared with the FACSCount. Conclusions: Evaluation of the image cytometer for CD4 and CD8 enumeration demonstrated comparable results with FCM on a population of HIV-1-infected patients. The image cytometer is a good alternative method for point-of-care settings in resource-constrained countries. © 2009 Clinical Cytometry Society.