Publication:
Low pressure leakage in anaesthetic machines evaluation by positive and negative pressure tests

dc.contributor.authorP. Somprakiten_US
dc.contributor.authorP. Soontrananen_US
dc.contributor.otherMahidol Universityen_US
dc.date.accessioned2018-07-04T07:31:31Z
dc.date.available2018-07-04T07:31:31Z
dc.date.issued1996-01-01en_US
dc.description.abstractPositive and negative pressure tests for checking the low pressure section of 36 anaesthetic machines in order to detect gas leaks were prospectively compared. The low pressure section was checked by the soap bubble test as a 'gold standard'. The leakage rate was 86.1%. The sensitivity and specificity of the positive pressure tests were 92% and 100% respectively, whilst the sensitivity and specificity of the negative pressure test were both 100%. However, seven machines revealed a leak which was only detected by the positive pressure test: this problem arose through defects in the emergency oxygen flush devices which allowed oxygen lo leak around the valve even though it was in the 'off' position. This might reduce the specificity of the negative pressure test to 41.7%. For security, both tests should be carried out in series; the positive pressure test followed by the negative pressure test. The possibility of gaseous leakage around the oxygen flush device is not generally recognised. This study emphasises the need for skilled servicing of anaesthetic machines on a regular basis. © 1996 The Association of Anaesthetists.en_US
dc.identifier.citationAnaesthesia. Vol.51, No.5 (1996), 461-464en_US
dc.identifier.doi10.1111/j.1365-2044.1996.tb07792.xen_US
dc.identifier.issn00032409en_US
dc.identifier.other2-s2.0-0029998913en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/17791
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=0029998913&origin=inwarden_US
dc.subjectMedicineen_US
dc.titleLow pressure leakage in anaesthetic machines evaluation by positive and negative pressure testsen_US