Publication: Personal respiratory protective devices: Efficacy of Millipore and Whatman filters
Issued Date
2000-02-01
Resource Type
ISSN
01252208
Other identifier(s)
2-s2.0-0034133092
Rights
Mahidol University
Rights Holder(s)
SCOPUS
Bibliographic Citation
Journal of the Medical Association of Thailand. Vol.83, No.2 (2000), 117-122
Suggested Citation
Kunchitthape Tanpowpong Personal respiratory protective devices: Efficacy of Millipore and Whatman filters. Journal of the Medical Association of Thailand. Vol.83, No.2 (2000), 117-122. Retrieved from: https://repository.li.mahidol.ac.th/handle/20.500.14594/26283
Research Projects
Organizational Units
Authors
Journal Issue
Thesis
Title
Personal respiratory protective devices: Efficacy of Millipore and Whatman filters
Author(s)
Other Contributor(s)
Abstract
A Millipore filter with 0.22 μm pore size and a Whatman grade 1 filter with > 11 μm particle retention were used to capture laser smoke particle mimic atmospheric suspended particulate matter. The experiment was conducted at the Department of Otolaryngology in Ramathibodi Hospital from April 1996 to October 1997. The laser smoke particle evacuator with rotameter created an air flow rate of 15 1/min through the filters. The mean and standard deviation of the laser smoke particle count under high power optical microscope in a 10 Millipore filter and a 10 Whatman filter were 411,327.6 +/-13,325.O and 290,453.0 +/-28,409.8 respectively, 29.4 per cent different. Laser smoke particle size distribution in both filters under eyepiece micrometer was: 1 to 10 μm in Millipore (99.0%) and in Whatman (96.2%), 1 to 5 μm in Millipore (77.1%) and in Whatman (77.6%), no laser smoke particle larger than 17 μm was detected. The Millipore filter ruptured when the air flow rate was greater than 15 1/min. The Whatman filter was suitable for evaluating filtration efficacy of various personal respiratory protective devices in a high air flow rate condition.