Publication:
Simultaneous colorimetric measurements of antioxidant capacity by flow injection analysis with paired emitter detector diode

dc.contributor.authorNantaya Manthongen_US
dc.contributor.authorPrapin Wilairaten_US
dc.contributor.authorDuangjai Nacaprichaen_US
dc.contributor.authorSumonmarn Chaneamen_US
dc.contributor.otherSilpakorn Universityen_US
dc.contributor.otherMahidol Universityen_US
dc.contributor.otherFlow Innovation-Research for Science and Technology Laboratories (FIRST Labs.en_US
dc.date.accessioned2020-01-27T08:16:28Z
dc.date.available2020-01-27T08:16:28Z
dc.date.issued2019-01-01en_US
dc.description.abstract© 2019, The Japan Society for Analytical Chemistry. An effective flow injection analysis (FIA) system employing paired emitter detector diode (PEDD) for simultaneous dual antioxidant assays is proposed. The total antioxidant capacity (TAC) was measured using the 2,2'-azino-bis(3- ethylbenzothiazoline-6-sulfonic acid) (ABTS) assay in parallel with the ferric reducing antioxidant power (FRAP) assay. Ascorbic acid was used as the reference antioxidant and the TAC value reported as ascorbic acid equivalent (AAE). A 10- port valve furnished with two injection loops allowed the sample from one loop to react with the ABTS reagent and be detected by one PEDD. At the same time, the sample from the second loop reacted with the FRAP reagent and is detected at the second PEDD. A pair of red light emitting diodes (634 nm) was employed for both PEDDs. The linearity range was 10 to 50 μM ascorbic acid, with limit of detection of 3.18 and 4.37 μM, and precision of 2.41 and 2.15% RSD (for 30 μM ascorbic acid, n = 10) for the ABTS and FRAP assay, respectively. Sample throughput of 90 samples/h was achieved. The method was applied to the measurement of TAC of commercial fruit juice, instant tea products and vitamin C tablets. The observed AAE values were in good agreement with those obtained using batch methods with a spectrophotometer.en_US
dc.identifier.citationAnalytical Sciences. Vol.35, No.5 (2019), 535-541en_US
dc.identifier.doi10.2116/analsci.18p512en_US
dc.identifier.issn13482246en_US
dc.identifier.issn09106340en_US
dc.identifier.other2-s2.0-85065514782en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/50591
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85065514782&origin=inwarden_US
dc.subjectChemistryen_US
dc.titleSimultaneous colorimetric measurements of antioxidant capacity by flow injection analysis with paired emitter detector diodeen_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85065514782&origin=inwarden_US

Files