Publication:
Radiation exposure in fluoroscopy of the shoulder: A comparison between mini and conventional c-arm image intensifier

dc.contributor.authorDirek Tantigateen_US
dc.contributor.authorKongkhet Riansuwanen_US
dc.contributor.authorWaiwit Sanguanwongwanen_US
dc.contributor.authorBanchong Mahaisavariyaen_US
dc.contributor.otherMahidol Universityen_US
dc.date.accessioned2018-11-09T02:43:13Z
dc.date.available2018-11-09T02:43:13Z
dc.date.issued2014-01-01en_US
dc.description.abstract© 2014, Medical Association of Thailand. All Rights Reserved. Objective: To compare the radiation exposure of the primary surgeon while using the conventional c-arm and mini c-arm fluoroscopy at the shoulder. Material and Method: Twelve shoulders of six fresh cadavers were used to simulate this experimental study. Radiation exposure of the primary surgeon was measured at the head and neck region. The average radiation dose was measured using both the conventional c-arm and mini c-arm, then the findings were compared. Results: Mean radiation energy of the conventional c-arm was significantly lower than the mini c-arm at 59.39±1.43 kV and 70.58±4.01 kV, respectively (p<0.001). Dose rates to which the primary surgeon was exposed from the conventional c-arm and the mini c-arm were 81.46±30.37 μSv/hour and 87.54±43.69 μSv/hour, respectively. However, the difference was not statistically significant (p = 0.875). Conclusion: There is no difference in the level of radiation safety for the primary surgeon when using the conventional c-arm vs. the mini c-arm for a fluoroscopic-assisted shoulder procedure. Therefore, selection of the methodology should be based on equipment availability and clinical considerations.en_US
dc.identifier.citationJournal of the Medical Association of Thailand. Vol.97, (2014), S88-S91en_US
dc.identifier.issn01252208en_US
dc.identifier.issn01252208en_US
dc.identifier.other2-s2.0-84922263325en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/34362
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84922263325&origin=inwarden_US
dc.subjectMedicineen_US
dc.titleRadiation exposure in fluoroscopy of the shoulder: A comparison between mini and conventional c-arm image intensifieren_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84922263325&origin=inwarden_US