Kanittha Kijsamanmith, Siriporn Timpawat, Noppakun Vongsavan, Bruce Matthews A comparison between red and infrared light for recording pulpal blood flow from human anterior teeth with a laser Doppler flow meter. Archives of Oral Biology. Vol.56, No.6 (2011), 614-618. doi:10.1016/j.archoralbio.2011.02.010 Retrieved from: https://repository.li.mahidol.ac.th/handle/20.500.14594/11540
Research Projects
Organizational Units
Authors
Journal Issue
Thesis
Title
A comparison between red and infrared light for recording pulpal blood flow from human anterior teeth with a laser Doppler flow meter
Other Contributor(s)
Abstract
Objective: To compare red (635 nm) and infrared (780 nm) light for recording pulpal blood flow from human anterior teeth with a laser Doppler flow meter. Design: Recordings were made from 7 healthy teeth in 5 subjects (aged 22-55 years) using a laser Doppler flow meter (Periflux 4001) equipped with both red and infrared lasers. Average blood flow signals were obtained with both light sources alternately from each tooth under five conditions: intact tooth without opaque rubber dam, intact tooth with dam, after injecting local anaesthetic (3% Mepivacaine) (LA) over the apex of the tooth and cavity preparation to almost expose the pulp, after removal and replacement of the pulp, and with the root canal empty. Results: With infrared light, because of technical limitations, data were obtained for the first three conditions only. The dam significantly decreased the mean blood flow by 82%. Injecting LA and cavity preparation had no significant effect. With red light, dam produced a decrease of 56%, and the resulting signal was reduced by 33% after LA and cavity preparation. The remaining signal fell by 46% after pulp removal and replacement. This contribution of the pulp is similar to that recorded previously with infrared light. There was no significant further change when the pulp was finally removed. Conclusions: The importance of using opaque rubber dam is confirmed. With dam, there is no advantage to using red rather than infrared light, and in each case the pulp contributes less than 50% to the blood flow signal. © 2010 Elsevier Ltd. All rights reserved.