Evaluation of image quality and lens's radiation dose of a low-dose cranial CT scan

dc.contributor.authorPipat Chiewviten_US
dc.contributor.authorJitladda Ananwattanasuken_US
dc.contributor.authorManus Mongkolsuken_US
dc.contributor.authorChulaluk Boonmaen_US
dc.contributor.authorSuthisak Suthipongchaien_US
dc.contributor.otherMahidol Universityen_US
dc.date.accessioned2018-09-13T06:59:44Z
dc.date.available2018-09-13T06:59:44Z
dc.date.issued2009-06-01en_US
dc.description.abstractObjective: To determine the lowest miliampere-second (mAs) of a cranial computed tomography (CT) scan that can maintain acceptable image quality on cranial CT scan which might help reducing the risk of cataract formation. Material and Method: The present study was performed on the 148 patients in routine daily practice sent for diagnosis of intracranial conditions by a cranial CT scan. During the cranial CT scanning, each lens's radiation dose on patient's eyes was measured by a thermoluminescent dosimeter. Clinical image quality, particularly in analysis of gray and white matter differentiation, was evaluated independently using a 5 point scale by two radiologists. Results: During standard cranial CT scan protocol with 250 miliampere-second (mAs), the total dose of left and right lens's dose were about 50.93 miligray (mGy) and 51.66 mGy, respectively. When applying low dose cranial CT scan protocols by decreasing mAs to 200, 150 and 100 mAs, the total dose of right and left lens were of 45.68 mGy and 46.04 mGy for 200 mAs, 34.65 mGy and 34.77 mGy for 150 mAs, 28.73 mGy and 29.25 mGy for 100 mAs respectively. Conclusion: A low dose cranial CT scan at 100 miliampere-second provides not only an acceptable clinical image quality, but also decreases the lens's radiation dose by 43%.en_US
dc.identifier.citationJournal of the Medical Association of Thailand. Vol.92, No.6 (2009), 831-835en_US
dc.identifier.issn01252208en_US
dc.identifier.issn01252208en_US
dc.identifier.other2-s2.0-67650444364en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/28054
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=67650444364&origin=inwarden_US
dc.subjectMedicineen_US
dc.titleEvaluation of image quality and lens's radiation dose of a low-dose cranial CT scanen_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=67650444364&origin=inwarden_US

Files

Collections