Publication:
Parametric study of pedicle screw: Effect of stress transfer parameters on pullout strength

dc.contributor.authorChompunut Somtuaen_US
dc.contributor.authorPanya Aroonjaratthamen_US
dc.contributor.authorKitti Aroonjaratthamen_US
dc.contributor.otherMahidol Universityen_US
dc.contributor.otherBurapha Universityen_US
dc.date.accessioned2022-08-04T07:59:44Z
dc.date.available2022-08-04T07:59:44Z
dc.date.issued2021-01-01en_US
dc.description.abstractA pedicle screw is an important implant used to fix and stabilize a cervical spine when spinal alignment is needed. The size and characteristics of pedicle screws affect pullout strength of the screws when inserted into the cervical spine. The pullout strength is greatly affected by the quality and duration of cervical spine treatment. To aid orthopedic surgeons in choosing a suitable pedicle screw for Thai cervical spine patients, the pullout strength of pedicle screws had to be considered. Since pedicle screws are different in design, size and characteristics, a middle ground parameter-Stress Transfer Parameter (STP) obtained through formula and finite element analysis, was used to compare the screws. Nine commonly-used pedicle screws were studied under the chosen parameters namely Outer Diameter (OD), Core Diameter (CD), Pitch (P), Proximal Root Radius (PRR), and Distal Root Radius (DRR). Each type of screw was inserted into C3 C4 C5 C6 and C7 of Thai cervical spines to bear the axial load in order to analyze their pullout conditions using Finite Element Analysis and to evaluate the STP. The maximum STP occurred on Y I pedicle screw as follows: 0.57 at C3, 0.48 at C4, 0.43 at C5, 0.48 at C6 and 0.59 at C7. The maximum STP resulted from the large OD and CD which could bear more axial load. The large Pitch decreased the stress concentration on each screw thread and the small PRR and DRR increased the slope of the blade to resist the pullout load.en_US
dc.identifier.citationAsia-Pacific Journal of Science and Technology. Vol.26, No.2 (2021)en_US
dc.identifier.doi10.14456/apst.2021.30en_US
dc.identifier.issn25396293en_US
dc.identifier.other2-s2.0-85123530357en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/123456789/75753
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85123530357&origin=inwarden_US
dc.subjectAgricultural and Biological Sciencesen_US
dc.subjectEngineeringen_US
dc.titleParametric study of pedicle screw: Effect of stress transfer parameters on pullout strengthen_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85123530357&origin=inwarden_US

Files