Publication:
Laboratory evaluation of the COBAS MIRA S random access analyser

dc.contributor.authorTaweesook Kanluanen_US
dc.contributor.authorSrisanit Intaramaneeen_US
dc.contributor.authorSurapon Tangvorasittichaien_US
dc.contributor.authorOrathai Tangvorasittichaien_US
dc.contributor.authorLerson Su Wantonen_US
dc.contributor.authorSudarat Manochiopinijen_US
dc.contributor.authorSoontharee Tantrarongrojen_US
dc.contributor.authorPimpan Leangphibulen_US
dc.contributor.otherMahidol Universityen_US
dc.date.accessioned2018-08-10T08:30:42Z
dc.date.available2018-08-10T08:30:42Z
dc.date.issued1991-01-01en_US
dc.description.abstractThe random access analyser COBAS MIRA S (Roche Diagnostics) was evaluated for two months. The instrument is a computer-controlled discrete analyser which can be run in a combination profile and/or single test mode. This instrument has special features, including an automatic cuvette segment changer, a reagent rack cooling system, an external keyboard and monitor, as well as a bar-code facility for the entry of test parameters, worklists and sample identification numbers. Study of within-run and between-run precision gave values of % CV = 0.54−3.37 and 0.61−3.65, respectively, for a variety of assays. Linearity testing to the upper limit of each test was also studied and were found to cover the necessary pathological range. Within the two-month period, no major problems were encountered. The instrument required minimum operator attention during operation. Correlation studies with the Hitachi 705 using six clinical chemistry tests (glucose, cholesterol, triglyceride, ALP, AST, ALT) gave correlation coefficients ranging from 0.95–0.99 and slopes of 0.91–1.17. © 1991, Taylor & Francis Ltd.en_US
dc.identifier.citationJournal of Automatic Chemistry. Vol.13, No.3 (1991), 97-100en_US
dc.identifier.doi10.1155/S1463924691000184en_US
dc.identifier.issn01420453en_US
dc.identifier.other2-s2.0-0025786886en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/21995
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=0025786886&origin=inwarden_US
dc.subjectBiochemistry, Genetics and Molecular Biologyen_US
dc.titleLaboratory evaluation of the COBAS MIRA S random access analyseren_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=0025786886&origin=inwarden_US