Publication:
Mechanical evaluation of stainless steel and titanium dynamic hip screws for trochanteric fracture

dc.contributor.authorSupakit Rooppakhunen_US
dc.contributor.authorNattapon Chantarapanichen_US
dc.contributor.authorBancha Chernchujiten_US
dc.contributor.authorBanchong Mahaisavariyaen_US
dc.contributor.authorSedthawatt Sucharitpwatskulen_US
dc.contributor.authorKriskrai Sitthiseripratipen_US
dc.contributor.otherSuranaree University of Technologyen_US
dc.contributor.otherPrince of Songkla Universityen_US
dc.contributor.otherFaculty of Medicine, Thammasat Universityen_US
dc.contributor.otherMahidol Universityen_US
dc.contributor.otherThailand National Metal and Materials Technology Centeren_US
dc.date.accessioned2018-09-24T08:59:52Z
dc.date.available2018-09-24T08:59:52Z
dc.date.issued2010-09-01en_US
dc.description.abstractThis study aimed to present the mechanical performance evaluation of the dynamic hip screw (DHS) for trochanteric fracture by means of finite element method. The analyses were performed based on stainless steel and titanium implant material definitions at various stages of bone healing and including implant removal. The assessment of the mechanical performance used two parameters, von Mises stress to evaluate the strength of bone and implant and elastic strain to evaluate fracture stability. The results show several critical aspects of dynamic hip screw for trochanteric fracture stabilization. In the initial stage of bone healing process, partial weight bearing should be applied to avoid the implant failure. In the late stage of bone healing, stainless steel implant should be removed.en_US
dc.identifier.citationWorld Academy of Science, Engineering and Technology. Vol.70, (2010), 662-665en_US
dc.identifier.issn20103778en_US
dc.identifier.issn2010376Xen_US
dc.identifier.other2-s2.0-78651544371en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/29077
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=78651544371&origin=inwarden_US
dc.subjectEngineeringen_US
dc.titleMechanical evaluation of stainless steel and titanium dynamic hip screws for trochanteric fractureen_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=78651544371&origin=inwarden_US

Files

Collections