Performance of the Access Bio/CareStart rapid diagnostic test for the detection of glucose-6-phosphate dehydrogenase deficiency: A systematic review and meta-analysis

dc.contributor.authorBenedikt Leyen_US
dc.contributor.authorAri Winasti Satyagrahaen_US
dc.contributor.authorHisni Rahmaten_US
dc.contributor.authorMichael E. von Frickenen_US
dc.contributor.authorNicholas M. Douglasen_US
dc.contributor.authorDaniel A. Pfefferen_US
dc.contributor.authorFe Espinoen_US
dc.contributor.authorLorenz von Seidleinen_US
dc.contributor.authorGisela Henriquesen_US
dc.contributor.authorNwe Nwe Ooen_US
dc.contributor.authorDidier Menarden_US
dc.contributor.authorSunil Parikhen_US
dc.contributor.authorGermana Banconeen_US
dc.contributor.authorAmalia Karahaliosen_US
dc.contributor.authorRic N. Priceen_US
dc.contributor.otherMelbourne School of Population and Global Healthen_US
dc.contributor.otherGokilaen_US
dc.contributor.otherEijkman Institute for Molecular Biologyen_US
dc.contributor.otherLondon School of Hygiene & Tropical Medicineen_US
dc.contributor.otherMenzies School of Health Researchen_US
dc.contributor.otherGeorge Mason University, Fairfax Campusen_US
dc.contributor.otherMahidol Universityen_US
dc.contributor.otherNuffield Department of Clinical Medicineen_US
dc.contributor.otherYale Universityen_US
dc.contributor.otherInstitut Pasteur, Parisen_US
dc.date.accessioned2020-01-27T09:20:40Z
dc.date.available2020-01-27T09:20:40Z
dc.date.issued2019-12-01en_US
dc.description.abstractBACKGROUND: To reduce the risk of drug-induced haemolysis, all patients should be tested for glucose-6-phosphate dehydrogenase (G6PD) deficiency (G6PDd) prior to prescribing primaquine (PQ)-based radical cure for the treatment of vivax malaria. This systematic review and individual patient meta-analysis assessed the utility of a qualitative lateral flow assay from Access Bio/CareStart (Somerset, NJ) (CareStart Screening test for G6PD deficiency) for the diagnosis of G6PDd compared to the gold standard spectrophotometry (International Prospective Register of Systematic Reviews [PROSPERO]: CRD42019110994). METHODS AND FINDINGS: Articles published on PubMed between 1 January 2011 and 27 September 2019 were screened. Articles reporting performance of the standard CSG from venous or capillary blood samples collected prospectively and considering spectrophotometry as gold standard (using kits from Trinity Biotech PLC, Wicklow, Ireland) were included. Authors of articles fulfilling the inclusion criteria were contacted to contribute anonymized individual data. Minimal data requested were sex of the participant, CSG result, spectrophotometry result in U/gHb, and haemoglobin (Hb) reading. The adjusted male median (AMM) was calculated per site and defined as 100% G6PD activity. G6PDd was defined as an enzyme activity of less than 30%. Pooled estimates for sensitivity and specificity, unconditional negative predictive value (NPV), positive likelihood ratio (LR+), and negative likelihood ratio (LR-) were calculated comparing CSG results to spectrophotometry using a random-effects bivariate model. Of 11 eligible published articles, individual data were available from 8 studies, 6 from Southeast Asia, 1 from Africa, and 1 from the Americas. A total of 5,815 individual participant data (IPD) were available, of which 5,777 results (99.3%) were considered for analysis, including data from 3,095 (53.6%) females. Overall, the CSG had a pooled sensitivity of 0.96 (95% CI 0.90-0.99) and a specificity of 0.95 (95% CI 0.92-0.96). When the prevalence of G6PDd was varied from 5% to 30%, the unconditional NPV was 0.99 (95% CI 0.94-1.00), with an LR+ and an LR- of 18.23 (95% CI 13.04-25.48) and 0.05 (95% CI 0.02-0.12), respectively. Performance was significantly better in males compared to females (p = 0.027) but did not differ significantly between samples collected from capillary or venous blood (p = 0.547). Limitations of the study include the lack of wide geographical representation of the included data and that the CSG results were generated under research conditions, and therefore may not reflect performance in routine settings. CONCLUSIONS: The CSG performed well at the 30% threshold. Its high NPV suggests that the test is suitable to guide PQ treatment, and the high LR+ and low LR- render the test suitable to confirm and exclude G6PDd. Further operational studies are needed to confirm the utility of the test in remote endemic settings.en_US
dc.identifier.citationPLoS medicine. Vol.16, No.12 (2019), e1002992en_US
dc.identifier.doi10.1371/journal.pmed.1002992en_US
dc.identifier.issn15491676en_US
dc.identifier.other2-s2.0-85076468070en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/123456789/51292
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85076468070&origin=inwarden_US
dc.subjectMedicineen_US
dc.titlePerformance of the Access Bio/CareStart rapid diagnostic test for the detection of glucose-6-phosphate dehydrogenase deficiency: A systematic review and meta-analysisen_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85076468070&origin=inwarden_US

Files

Collections