Publication:
Accuracy of transcutaneous bilirubin measured by the BiliCare device in late preterm and term neonates

dc.contributor.authorRatchada Kitsommarten_US
dc.contributor.authorBuranee Yangtharaen_US
dc.contributor.authorPunnanee Wutthigateen_US
dc.contributor.authorBosco Paesen_US
dc.contributor.otherMahidol Universityen_US
dc.contributor.otherMcMaster University, Faculty of Health Sciencesen_US
dc.date.accessioned2018-12-11T03:17:16Z
dc.date.accessioned2019-03-14T08:01:56Z
dc.date.available2018-12-11T03:17:16Z
dc.date.available2019-03-14T08:01:56Z
dc.date.issued2016-11-16en_US
dc.description.abstract© 2016 Taylor & Francis. Objective: The objective of this study is to explore the accuracy and performance of a new transcutaneous bilirubinometer (TCB) for the screening of jaundice in late preterm and term infants. Methods: A cross-sectional study was conducted. TCB measurements were performed using the BiliCareTMbilirubinometer. Paired TCB and serum bilirubin (SB) measurements were analysed. Results: One hundred and fourteen paired samples were collected from 93 healthy late preterm and term infants. Bilirubin measurements were done at median (interquartile range) of 50.5 (34, 72) hours. The mean (SD) difference between the TCB and SB was 1.87 (1.98) mg/dL. A TCB cut-off level at 8.0 mg/dL provides a sensitivity of 97.3% with a negative predictive value (NPV) of 87.5% to detect a SB level of at least 8.0 mg/dL. For SB levels of at least 10.0 mg/dL, a TCB cut-off at 9.0 mg/dL shows a sensitivity of 97.5%; NPV 95.4%. For a SB level of at least 13.0 mg/dL, a TCB cut-off at 12 or 13 mg/dL had a sensitivity of 92.9% and NPV of 98.7%. Conclusion: The BiliCareTMdemonstrated good performance with positive bias for the screening of jaundice in healthy late preterm or term infants. However, if adopted, proper cut-off levels should be chosen because of sub-optimal device precision.en_US
dc.identifier.citationJournal of Maternal-Fetal and Neonatal Medicine. Vol.29, No.22 (2016), 3641-3645en_US
dc.identifier.doi10.3109/14767058.2016.1140141en_US
dc.identifier.issn14764954en_US
dc.identifier.issn14767058en_US
dc.identifier.other2-s2.0-84958038440en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/123456789/40998
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84958038440&origin=inwarden_US
dc.subjectMedicineen_US
dc.titleAccuracy of transcutaneous bilirubin measured by the BiliCare device in late preterm and term neonatesen_US
dc.typeArticleen_US
dspace.entity.typePublication
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84958038440&origin=inwarden_US