Label-free Electrochemical Immunosensor Toward IGFBP-7 Biomarker for Early Monitoring of Acute Kidney Injury
| dc.contributor.author | Unson S. | |
| dc.contributor.author | Ruankham W. | |
| dc.contributor.author | Phopin K. | |
| dc.contributor.author | Tantimongcolwat T. | |
| dc.contributor.correspondence | Unson S. | |
| dc.contributor.other | Mahidol University | |
| dc.date.accessioned | 2026-03-27T18:08:19Z | |
| dc.date.available | 2026-03-27T18:08:19Z | |
| dc.date.issued | 2026-01-01 | |
| dc.description.abstract | Insulin-like growth factor-binding protein 7 (IGFBP-7) is a potential biomarker for early diagnosis of acute kidney injury (AKI). Herein, we have developed a novel label-free electrochemical immunosensor based on the self-assembly of 6-(ferrocenyl)hexanethiol (Fc) and antibody-coupled L-cysteine on the gold nanoparticle-modified electrode (Ab/Fc-Au/SPCE) for sensitive detection of IGFBP-7 in urine. Fc was directly embedded on the electrode surface to serve as an intrinsic redox signal for the IGFBP-7 immunosensor. The binding of IGFBP-7 target to the immobilized antibody resulted in a decrease in the peak current of the differential pulse voltammetry (DPV) measurement. The Ab/Fc-Au/SPCE exhibited a good linear DPV response to IGFBP-7 concentrations ranging from 0.0001 to 100.0 pg mL<sup>-1</sup>, with a limit of detection (LOD) of 0.07 pg mL<sup>-1</sup>. Validation using spiked artificial urines showed recovery rates in a range of 79.8 to 109.3%. The proposed immunosensor demonstrated good specificity and long-term stability, which are crucial properties for biosensor implementation. Importantly, the electrochemical IGFBP-7 immunosensor can function without the need for external redox agents, requiring only simple steps of sample incubation and subsequent PBS buffer rinsing. Our findings demonstrate a promising potential of Ab/Fc-Au/SPCE immunosensor as a Point-of-Care Testing (POCT) device for rapid detection of IGFBP-7 in early AKI diagnosis. | |
| dc.identifier.citation | IEEE Sensors Journal (2026) | |
| dc.identifier.doi | 10.1109/JSEN.2026.3671274 | |
| dc.identifier.eissn | 15581748 | |
| dc.identifier.issn | 1530437X | |
| dc.identifier.scopus | 2-s2.0-105032879938 | |
| dc.identifier.uri | https://repository.li.mahidol.ac.th/handle/123456789/115838 | |
| dc.rights.holder | SCOPUS | |
| dc.subject | Physics and Astronomy | |
| dc.subject | Engineering | |
| dc.title | Label-free Electrochemical Immunosensor Toward IGFBP-7 Biomarker for Early Monitoring of Acute Kidney Injury | |
| dc.type | Article | |
| mu.datasource.scopus | https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=105032879938&origin=inward | |
| oaire.citation.title | IEEE Sensors Journal | |
| oairecerif.author.affiliation | Mahidol University |