The effect of the AN69 ST100 versus oXiris<sup>®</sup> membrane on filter life during CRRT for non-anticoagulated liver failure patients.

dc.contributor.authorNaorungroj T.
dc.contributor.authorBaldwin I.
dc.contributor.authorEastwood G.
dc.contributor.authorBellomo R.
dc.contributor.otherMahidol University
dc.date.accessioned2023-05-19T07:36:57Z
dc.date.available2023-05-19T07:36:57Z
dc.date.issued2023-04-01
dc.description.abstractBackgrounds: In patients with liver failure, continuous renal replacement therapy (CRRT) without anticoagulation may be necessary. A new heparin coated membrane (oXiris®) may prolong circuit life in this setting. Objectives: In liver failure patients not receiving anticoagulation, to compare CRRT circuit life with the oXiris® versus the AN69 ST100 (usual care) membrane. Methods: Randomized single cross-over trial. Results: We studied 20 patients and 39 circuits. Twenty-five treatments used femoral and 14 internal jugular access catheters. Median circuit life was 21 h (IQR: 8.25–35.5) with the AN69 versus 16.0 h (14–25) with the oXiris® membrane (p = 0.36). Median first circuit duration was 14 (11.25–23) h for the AN69 ST100 versus 16 (8 –26) for the oXiris® membrane. There was also no difference between the AN69 ST100 or oXiris® membrane circuits using femoral access at 13 (8 –22.5) versus 15.5 (12.5–21.5) h (p = 0.57) or internal jugular access at 28 (13–47) versus 23 (21–29) h (p = 0.79), respectively. Conclusions: The oXiris® heparin-grafted membrane does not appear to prolong circuit life in liver failure patients receiving CRRT without anticoagulation.
dc.identifier.citationInternational Journal of Artificial Organs Vol.46 No.4 (2023) , 209-214
dc.identifier.doi10.1177/03913988231155253
dc.identifier.eissn17246040
dc.identifier.issn03913988
dc.identifier.pmid36895128
dc.identifier.scopus2-s2.0-85150615486
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/81700
dc.rights.holderSCOPUS
dc.subjectChemical Engineering
dc.titleThe effect of the AN69 ST100 versus oXiris<sup>®</sup> membrane on filter life during CRRT for non-anticoagulated liver failure patients.
dc.typeArticle
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85150615486&origin=inward
oaire.citation.endPage214
oaire.citation.issue4
oaire.citation.startPage209
oaire.citation.titleInternational Journal of Artificial Organs
oaire.citation.volume46
oairecerif.author.affiliationSiriraj Hospital
oairecerif.author.affiliationUniversity of Melbourne
oairecerif.author.affiliationMonash University
oairecerif.author.affiliationRoyal Melbourne Hospital
oairecerif.author.affiliationAustin Hospital
oairecerif.author.affiliationANZIC Research Centre

Files