Evaluation of the Efficacy and Safety of the ITM <sup>68</sup>Ge/<sup>68</sup>Ga Generator After its Recommended Shelf-life

dc.contributor.authorSriprapa T.
dc.contributor.authorDoungta T.
dc.contributor.authorSakulsamart N.
dc.contributor.authorTaweewatthanasopon N.
dc.contributor.authorMadputeh L.
dc.contributor.authorRagchana P.
dc.contributor.authorSritongkul N.
dc.contributor.authorTantawiroon M.
dc.contributor.authorKongmuang S.
dc.contributor.authorKhiewvan B.
dc.contributor.authorShiratori S.
dc.contributor.otherMahidol University
dc.date.accessioned2023-10-31T18:02:03Z
dc.date.available2023-10-31T18:02:03Z
dc.date.issued2023-01-01
dc.description.abstractObjective:68Ga can be routinely produced by a 68Ge/68Ga generator without the need for a cyclotron. It is recommended to replace the 68Ge/68Ga generator after 250 elutions or 12 months of shelf-life whichever endpoint is reached first. However, a 68Ge/68Ga generator that has gone past its recommended lifespan can still be further used as a 68Ga source for 68Ga-labeled radiopharmaceuticals for use in animal experiments. To ensure the quality of 68Ga eluates, we aimed to evaluate the efficacy and safety of the ITM (Isotope Technologies München) 68Ge/68Ga generator in our institute after its recommended shelf-life. Materials and Methods: A 21-month-old ITM 68Ge/68Ga generator was eluted using 4.0 ml of 0.05 M HCl. The 68Ga elution yields were calculated, and 68Ge breakthrough was measured at least 48 h after elution in an aliquot amount using a multichannel analyzer (MCA) with a high-purity germanium probe. Metal impurities in the 68Ga eluates were analyzed by ICP-MS. Results: The elution yield of 68Ga was 35.2 ± 8.1%; n = 5 (decay corrected). 68Ge breakthrough from the ITM 68Ge/68Ga generator was below the detectable level. The average amounts of the metallic ions 57Fe, 66Zn, 203Pb, 60Ni, and 63Cu were 18.60, 9.86, 2.42, 0.52, and 0.47 μg/GBq, respectively. Conclusion: The ITM 68Ge/68Ga generator demonstrated consistent and reliable 68Ga elution profiles with an absence of either 68Ge breakthrough or other metal contaminants in the eluent samples as verified by the manufacturer. The use of the ITM 68Ge/68Ga generator could be extended past its recommended shelf-life to prepare 68Ga radiopharmaceuticals that are considered safe and suitable for use in animal experimentation and other applications.
dc.identifier.citationSiriraj Medical Journal Vol.75 No.10 (2023) , 752-758
dc.identifier.doi10.33192/smj.v75i10.264289
dc.identifier.eissn22288082
dc.identifier.scopus2-s2.0-85174724718
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/123456789/90872
dc.rights.holderSCOPUS
dc.subjectMedicine
dc.titleEvaluation of the Efficacy and Safety of the ITM <sup>68</sup>Ge/<sup>68</sup>Ga Generator After its Recommended Shelf-life
dc.typeArticle
mu.datasource.scopushttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85174724718&origin=inward
oaire.citation.endPage758
oaire.citation.issue10
oaire.citation.startPage752
oaire.citation.titleSiriraj Medical Journal
oaire.citation.volume75
oairecerif.author.affiliationSiriraj Hospital
oairecerif.author.affiliationThailand Institute of Nuclear Technology, Bangkok
oairecerif.author.affiliationSilpakorn University
oairecerif.author.affiliationMahidol University

Files