Determination of cystatin C using paper-based analytical devices for early detection of renal failure
dc.contributor.author | Sabarudin A. | |
dc.contributor.author | Tyas A.A. | |
dc.contributor.author | Pamungkas B.D. | |
dc.contributor.author | Pratama I.G.B.A. | |
dc.contributor.author | Andayani U. | |
dc.contributor.author | Sakti S.P. | |
dc.contributor.other | Mahidol University | |
dc.date.accessioned | 2023-11-27T18:03:00Z | |
dc.date.available | 2023-11-27T18:03:00Z | |
dc.date.issued | 2023-08-21 | |
dc.description.abstract | Renal failure is often detected when it has reached the last stage. The amount of cystatin C in blood serum can be used as a biomarker for early identification of renal function failure. This work reported a fast and environmentally friendly microfluidic paper-based analytical device (μPAD) to determine the amount of cystatin C in the artificial blood serum samples. Functionalized gold nanoparticles (AuNPs) were prepared by capping the surface of AuNPs with papain to form AuNPs-papain. The aggregate of functionalized AuNPs-papain and Cu2+ ion could be formed due to coordination interaction between Cu2+ ion and papain attached to AuNPs. In the presence of cystatin C, the amount of aggregate decreased because the functionalized AuNPs-papain would bind easily with cystatin C as the substrate of papain. The intensity of aggregate on μPAD was scanned using a paper scanner. The difference intensity of aggregate in the presence of cystatin C was proportional to the concentration of cystatin C in the artificial blood serum sample. The linearity range obtained by this work was 0.5-10.0?mg/L with regression of y=0.3302x+3.488 and linearity of r2=0.998. The limit of detection (LOD) and limit of quantitation (LOQ) were 0.511 (3xSD/slope) and 2.044?mg/L (12xSD/slope), respectively. The linearity range and the LOD were wide enough to detect cystatin C in normal people or people with renal diseases. | |
dc.identifier.citation | AIP Conference Proceedings Vol.2818 No.1 (2023) | |
dc.identifier.doi | 10.1063/5.0131190 | |
dc.identifier.eissn | 15517616 | |
dc.identifier.issn | 0094243X | |
dc.identifier.scopus | 2-s2.0-85176953660 | |
dc.identifier.uri | https://repository.li.mahidol.ac.th/handle/20.500.14594/91213 | |
dc.rights.holder | SCOPUS | |
dc.subject | Physics and Astronomy | |
dc.title | Determination of cystatin C using paper-based analytical devices for early detection of renal failure | |
dc.type | Conference Paper | |
mu.datasource.scopus | https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85176953660&origin=inward | |
oaire.citation.issue | 1 | |
oaire.citation.title | AIP Conference Proceedings | |
oaire.citation.volume | 2818 | |
oairecerif.author.affiliation | Brawijaya University | |
oairecerif.author.affiliation | Mahidol University |