Publication:
Comparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selection

dc.contributor.authorSabong Srivannaboonen_US
dc.contributor.authorChareenun Chirapapaisanen_US
dc.contributor.authorPratuangsri Chonpimaien_US
dc.contributor.authorSunisa Koodkaewen_US
dc.contributor.otherMahidol Universityen_US
dc.date.accessioned2018-11-23T11:01:46Z
dc.date.available2018-11-23T11:01:46Z
dc.date.issued2015-01-01en_US
dc.description.abstract© 2015 ASCRS and ESCRS. Purpose To compare corneal astigmatism measurements obtained using a new optical biometer (AL-Scan) and a standard biometer based on partial coherence interferometry (IOLMaster 500) for toric intraocular lens (IOL) selection. Setting Siriraj Hospital, Mahidol University, Bangkok, Thailand. Design Prospective comparative study. Methods Keratometry (K) readings were taken with the 2 biometers in eyes of cataract patients. The corneal astigmatism readings from the 2 devices were compared. The cylinder measurements were transformed into J0 and J45 vector components for analysis, and the devices' values were compared using the paired t test. Correlation coefficients were calculated. The Bland-Altman method was used to evaluate the agreement between the sets of measurements. Results One hundred thirty-seven eyes (81 patients) were enrolled. The magnitude of astigmatism between the new biometer 2.4 mm zone and the standard biometer was similar (P =.19). There was a statistically significant difference between values from the new biometer 3.3 mm zone and the standard biometer (P <.05). There was good correlation in the overall magnitude of astigmatism between devices. There was a trend of low correlation toward the flat cornea (K <43.0 diopters [D]), low corneal cylinder (<1.0 D), and with-the-rule astigmatism. The IOL toricity calculated from both devices was similar. Conclusions Corneal astigmatism measurements with the 2 biometers were similar in magnitude, especially in 2.4 mm zone of the new biometer. There was no difference in the toric IOL selections between the devices. Financial Disclosure No author has a financial or proprietary interest in any material or method mentioned.en_US
dc.identifier.citationJournal of Cataract and Refractive Surgery. Vol.41, No.2 (2015), 364-371en_US
dc.identifier.doi10.1016/j.jcrs.2014.06.029en_US
dc.identifier.issn18734502en_US
dc.identifier.issn08863350en_US
dc.identifier.other2-s2.0-84923095888en_US
dc.identifier.urihttps://repository.li.mahidol.ac.th/handle/20.500.14594/36757
dc.rightsMahidol Universityen_US
dc.rights.holderSCOPUSen_US
dc.source.urihttps://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84923095888&origin=inwarden_US
dc.subjectMedicineen_US
dc.titleComparison of corneal astigmatism measurements of 2 optical biometer models for toric intraocular lens selectionen_US
dc.typeArticleen_US