Publication: Performance evaluation of the automated nucleated red blood cell enumeration on Sysmex XN analyser
Issued Date
2015-06-01
Resource Type
ISSN
1751553X
17515521
17515521
Other identifier(s)
2-s2.0-84928400833
Rights
Mahidol University
Rights Holder(s)
SCOPUS
Bibliographic Citation
International Journal of Laboratory Hematology. Vol.37, No.3 (2015), 341-345
Suggested Citation
C. Tantanate, C. Klinbua Performance evaluation of the automated nucleated red blood cell enumeration on Sysmex XN analyser. International Journal of Laboratory Hematology. Vol.37, No.3 (2015), 341-345. doi:10.1111/ijlh.12291 Retrieved from: https://repository.li.mahidol.ac.th/handle/20.500.14594/35443
Research Projects
Organizational Units
Authors
Journal Issue
Thesis
Title
Performance evaluation of the automated nucleated red blood cell enumeration on Sysmex XN analyser
Author(s)
Other Contributor(s)
Abstract
© 2014 John Wiley & Sons Ltd. Introduction: Presence of peripheral blood nucleated red blood cell (NRBC) is associated with pathological conditions and leads to the overestimation of white blood cell count in automated haematology analysers (HA). The authors evaluated NRBC enumeration by a new HA Sysmex XN (XN) to demonstrate the precision and comparability to manual count (MC) at the various NRBC values. Methods: Specimens with initially NRBC positive were included. For precision assessment, 8 levels of NRBCs were repeatedly analysed. For comparison study, 234 specimens were analysed by both XN and MC. Results: For precision study, the percentage of coefficient of variation ranged from 14% to 45.6% and 1.2% to 4.4% for MC and XN, respectively. For comparison study between XN and MC, NRBCs ranged from 0% to 612.5%. Regression analysis demonstrated an r2of 0.98. The mean bias of 14.1% with 95% limits of agreement between -48.76% and 76.95% was found. The NRBC counts from XN appeared to be more in accordance with MC when the NRBCs were lower than 200% with the concordance rate of 94.2%. Conclusion: The automated NRBC enumeration by XN was precise and could replace the traditional MC, especially for the specimens with NRBCs lower than 200%.