Publication: Indirect determination of mercury by inhibition of glucose oxidase immobilized on a carbon paste electrode
Issued Date
2012-02-01
Resource Type
ISSN
14523981
Other identifier(s)
2-s2.0-84857880276
Rights
Mahidol University
Rights Holder(s)
SCOPUS
Bibliographic Citation
International Journal of Electrochemical Science. Vol.7, No.2 (2012), 1001-1010
Suggested Citation
A. Samphao, H. Rerkchai, J. Jitcharoen, D. Nacapricha, K. Kalcher Indirect determination of mercury by inhibition of glucose oxidase immobilized on a carbon paste electrode. International Journal of Electrochemical Science. Vol.7, No.2 (2012), 1001-1010. Retrieved from: https://repository.li.mahidol.ac.th/handle/123456789/13987
Research Projects
Organizational Units
Authors
Journal Issue
Thesis
Title
Indirect determination of mercury by inhibition of glucose oxidase immobilized on a carbon paste electrode
Other Contributor(s)
Abstract
An amperometric biosensor for the indirect determination of mercury was developed by inhibition of glucose oxidase immobilized on a carbon paste electrode bulk modified with manganese dioxide 5% (m:m). Optimum conditions for biosensor were found as: operating potential 0.46 V vs Ag/AgCl, enzyme load 0.56 U/mm 2 electrode surface, supporting electrolyte: 0.1 M citrate phosphate buffer (pH 7). The current response of the biosensor in hydrodynamic amperometry was linearity dependent on the concentration within 2.0 - 32.5 mg/L Hg (II) with a detection limit of 0.5 mg/L. The enzymatic activity could be fully recovered by exposure of the electrode to a solution of EDTA (0.1 M) for 2 minutes. Minor interferences were caused by Cr 3+ and Zn 2+ . Analytical results of spiked water sample showed a good correlation with determinations by atomic absorption spectrometry where the difference is not statistically significant at 95% confidence level. The sensor yielded a stable and reproducible response more than 2 weeks when stored dry at 4 °C. © 2012 by ESG.