Publication: Optimization of acquisition parameters for simultaneous 201Tl and 99mTc dual-isotope myocardial imaging
Submitted Date
Received Date
Accepted Date
Issued Date
2005-10-01
Copyright Date
Announcement No.
Application No.
Patent No.
Valid Date
Resource Type
Edition
Resource Version
Language
File Type
No. of Pages/File Size
ISBN
ISSN
00189499
eISSN
Scopus ID
WOS ID
Pubmed ID
arXiv ID
Call No.
Other identifier(s)
2-s2.0-29144504393
Journal Title
Volume
Issue
item.page.oaire.edition
Start Page
End Page
Access Rights
Access Status
Rights
Mahidol University
Rights Holder(s)
SCOPUS
Physical Location
Bibliographic Citation
IEEE Transactions on Nuclear Science. Vol.52, No.5 I (2005), 1227-1235
Citation
Wen Tung Wang, Benjamin M W Tsui, David S. Lalush, Chiraporn Tocharoenchai, Eric C. Frey (2005). Optimization of acquisition parameters for simultaneous 201Tl and 99mTc dual-isotope myocardial imaging. Retrieved from: https://repository.li.mahidol.ac.th/handle/123456789/16494.
Research Projects
Organizational Units
Authors
Journal Issue
Thesis
Title
Optimization of acquisition parameters for simultaneous 201Tl and 99mTc dual-isotope myocardial imaging
Alternative Title(s)
Author's Affiliation
Author's E-mail
Editor(s)
Editor's Affiliation
Corresponding Author(s)
Creator(s)
Compiler
Advisor(s)
Illustrator(s)
Applicant(s)
Inventor(s)
Issuer
Assignee
Series
Has Part
Abstract
In201Tl/99mTc dual-isotope simultaneous-acquisition (DISA) myocardial imaging, crosstalk due to Tc photons results in significant contamination of the Tl data. The objective of this work is to seek the acquisition parameters (i.e., energy window width and center) that have the optimal tradeoff between minimizing the crosstalk and maximizing the detection efficiency. The optimization criterion was based on maximizing an ideal observer signal-to-noise ratio (SNR) for the myocardial defect detection task using single-isotope and DISA projection images acquired from a torso phantom. For single-isotope images, the optimal energy windows (width/center: 26 keV/75 keV and 28 keV/165 keV for201Tl, 30 keV/142 keV for99mTc) are wider than typical windows. For DISA imaging, the optimal windows varied with the99mTc to201Tl activity ratio and are thus likely to depend on the uptake ratio in each patient. Using the optimal ratio 2.25-2.75 (148 MBq201Tl and 333-407 MBq99mTc) with the corresponding optimal windows (22 keV/72 keV, 24 keV/167 keV, and 24 keV/140 keV) gives201Tl images with substantially increased SNRs as well as99mTc images with SNRs same as those of 370 MBq99mTc-only images. However, without the addition of crosstalk compensation, the use of the optimal activity and energy windows alone is likely not sufficient to restore the DISA Tl SNR to that of Tl-only image. © 2005 IEEE.